| HEPA filter leak detection >> Automatic filter detection device >> TDA-100P 自动式过滤器检测装置 |  |
 |
| |  | Product Id: | 2914193916 | Product Name: | TDA-100P 自动式过滤器检测装置 | Spec: | TDA-100P | Product Memo: | made in USA | Product Class: | HEPA filter leak detection | | |
|
Product Content |
Product Model:: TDA-100P Product Name: automatic filter detection device - This is a test using the most advanced technology by the microcomputer control processor and automatic filter detection device, in order to debug a user-friendly operation, provides a friendly interface - all functions and movements of the test parameters are available through 4 set of function keys. Collection efficiency. Filter pressure loss and filter performance test flow test parameters, etc. can be directly displayed in the respective figures show panel. Through micro-computer processor to monitor these three parameters, the parameter exceeds a set point, PAIL bright lights, the parameters for the digital display flashes and alarm reset before the no testing - the device can be easily installed in the user - prepared to operate in Taichung, and through the RS-232C and computer connected to the data taken and control. . Product Content Specifications each detail and each part.
Particles occurred in principle and substance occurred Sprayers: DOP, PAO Size 0.18 Ucmd: the basis and foundation of Japan’s national 42CFRPart84 standards Geometric standard deviation <1.6: According to 42CFRPart84 and a basis for Japan’s national standard for the certification of Mass concentration 20 ~ 150mg/m3
● planktonic particles test
Test particle Scattered near the front Test range 0.0001-200mg/m3 Testing accuracy ± 1% Flow 10-100L/min
● Stress Test
Test particle Electronic pressure converter Test range 0-100H2O Testing accuracy Full range of ± 0.5%
● efficiency test
Test flow 10-100L/min Test range 0-99.9995%
● Sampling Flow Testing
Test the principle of mass flowmeter Test range 10-100L/min Repeatability ± 0.5%
● Other
Power Supply 115V AC, 60Hz, 2A Output RS-232C The need for compressed air source 339.6L/min or 5.6Kg/m3 Dimensions About 534 × 813 × 788mm
** Use 20psi (0.14Mpa) pressure Note: do not recommend that DOP |
|
 |